Background

With the MiniFlex, it is possible to select Soller slits with small widths in order to obtain high resolution. In particular, the 0.5° receiving Soller slit and the 5 mm or 2 mm incident high limiting (DHL) slits can suppress the axial divergence, or umbrella effect which is prominent on the low-angle side. This report shows a comparison of the X-ray diffraction profiles of zeolite-A when the receiving Soller slit and DHL are changed (the incident Soller slit is 2.5°).